Chapter 2
Using the Module
2-12
ni.com
To convert module readings to strain use the following equation:
Full-Bridge Type I
This section provides information for the full-bridge strain-gage configuration type I. The
full-bridge type I only measures bending strain. Figure 2-11 shows how to position
strain-gage resistors in a bending configuration. Figure 2-12 shows the full-bridge type I
circuit wiring diagram.
Figure 2-11.
Full-Bridge Type I Rejecting Axial and Measuring Bending Strain
A full-bridge type I configuration has the following characteristics:
•
Four active strain-gage elements. Two strain-gage elements are mounted in the direction
of bending strain on one side of the strain specimen (top) while the other two are mounted
in the direction of bending strain on the opposite side (bottom).
•
Highly sensitive to bending strain.
•
Rejects axial strain.
•
Sensitivity ~ 2
μ
V/V per
με
, for GF = 2.0.
strain
(
ε )
2
V
r
–
GF
------------
=
R
2
Rejects Axial
Measures Bending
(+ )
(– )
(+ )
(– )
R
1
R
3
R
4
R
3
R
4
R
2
R
1