Chapter 2
Using the Module
©
National Instruments Corporation
2-3
Figure 2-2.
Connecting Remote Sense Wires to the NI PXIe-4330/4331
The actual bridge excitation voltage measured at the bridge is smaller than the voltage sourced
at the EX+ and EX– connector pins. This reduction in voltage is due to the voltage drop across
the excitation lead wire resistance (
R
lead
). If you do not use remote sensing of the actual bridge
voltage, the resulting reduction in gain is given by the following equations:
•
for half-bridge sensors
•
for full-bridge sensors
If you connect the remote sense signals directly to the bridge resistors, the
NI PXIe-4330/4331 senses the actual bridge voltage using high-impedance RS leads and
eliminates the gain errors caused by the resistance of the EX+ and EX– leads.
EX+
RS+
RS–
EX–
AI–
AI+
R
g
R
g
R
g
R
g
PXIe-4330/4331
R
lead
R
lead
R
lead
R
g
------------
2
R
lead
⋅
R
g
-------------------