Chapter 2
Using the Module
©
National Instruments Corporation
2-11
Figure 2-10.
Half-Bridge Type II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•
R
1
and R
2
—Half-bridge completion resistors located inside the NI PXIe-4330/4331.
•
R
3
—Active element measuring compressive strain (–
ε
).
•
R
s
—Shunt calibration resistor located inside the NI PXIe-4330/4331.
•
R
4
—Active element measuring tensile strain (+
ε
).
•
R
L
—Lead resistance.
•
GF
—Gage Factor, specified by the gage manufacturer.
•
V
CH
—Measured voltage of the bridge.
•
V
EX
—Excitation voltage provided by the NI PXIe-4330/4331.
V
r
—Offset compensated ratiometric bridge output defined by the following equation:
Note
The ratio of the bridge output voltage and the excitation voltage is done internally
on the NI PXIe-4330/4331.
+
–
+
–
AI+
EX+
EX–
R
L
R
L
R
L
RS–
QTR
/SC
Shunt
Cal A
+
–
PXIe-4330/4331 Set Bridge
Configuration to Half Bridge
TB-4330
Transducer
V
CH
R
1
R
2
R
4
(gage)
V
EX
R
3
(gage)
R
L
R
L
RS+
R
L
R
S
V
r
V
CH
strained
(
)
V
CH
unstrained
(
)
–
V
EX
---------------------------------------------------------------------------------
=