PXIe-4309 and TB-4309 (ST)/TB-4309 (MT)
|
© National Instruments
|
7
Figure 3.
TB-4309 (ST) Parts Locator Diagram
Caution
Make sure no exposed wire extends past the screw terminal. Exposed wire
increases the risk of a short circuit that can cause circuit failure.
5.
Insert the stripped end of the shielded signal wires into the appropriate terminal.
Note
Refer to the label next to each screw terminal to determine the function of the
terminal.
Note
Refer to the
PXIe-4309 User Manual
for wiring diagrams showing how to
connect signals to the TB-4309 (ST)/TB-4309 (MT).
6.
Tighten the terminal screws to a torque of 0.2 to 0.25 N · m (1.77 to 2.21 lb · in.).
Caution
Any wires connected to ground must be sufficiently insulated from any
voltage present.
1
Captive Strain-Relief Screw
2
Strain-Relief Bar
3
Tie Wrap Holes
4
Chassis Ground Screw
5
Terminal Block to Module Connector
6
Captive Top Cover Screw
7
Top Cover
8
Tie Wraps
9
Shielded Signal Wires
3
1
2
1
4
6
7
6
5
4
8
9
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com