Table 15. Simultaneous Sampled Analog Input Signal Source Selection (Continued)
Input 1 (Inverters)
Input 2 (AI Connector)
GPIC Input
Name
Switch
Contact Pair
Name
Switch
Contact Pair
AI9±
U DC 2
SW35
3, 4
ssAI9±
SW35
1, 2
AI10±
U DC 3
SW37
3, 4
ssAI10±
SW37
1, 2
AI11±
U DC 4
SW43
3, 4
ssAI11±
SW43
1, 2
AI12±
U DC 5
SW39
3, 4
ssAI12±
SW39
1, 2
AI13±
U DC 6
SW24
3, 4
ssAI13±
SW24
1, 2
AI14±
—
—
—
ssAI14±
connected
—
AI15±
—
—
—
ssAI15±
connected
—
Notice
Only one signal source should be connected to each GPIC analog input at
one time. Connecting two signal sources in parallel can produce unexpected results.
Analog inputs from inverters are shared between the half-bridge modules and the full-bridge
modules. For proper operation, each analog line should be driven by only one inverter. Refer
to
Inverter Signals
on page 9 for more information.
Low Pass Filters
The sbRIO-9687 provides a low pass filter for each analog input. The filter has two stages that
are populated by default with 10 kΩ resistors and 3.3 nF capacitors. Refer to the
sbRIO-9687
Specifications
for details about the filter characteristics.
Figure 10. Simultaneous Sampled AI Input Low Pass Filter
From Input
Selection
To Gain
Stage
C1
C2
R1
R2
R1
R2
The first two filter resistors (R1) have a second functionality on the board and have a value of
10 kΩ. Other filter elements can be changed to set filter characteristics as required by the
application. The R2 has a default value of 10 kΩ (0603 footprint), and C1 and C2 have a
default value of 3.3 nF (0805 footprint).
sbRIO-9687 User Manual
|
© National Instruments
|
21