![National Instruments sbRIO-9218 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/national-instruments/sbrio-9218/sbrio-9218_getting-started-manual_3684734007.webp)
Full-Bridge Signal Conditioning
+ –
Reference
Input
AI+
EX+
RS+
AI–
EX–
RS–
SC
SC
sbRIO-9218
ADC
• The analog input connections sense then amplify the incoming analog signal.
• The excitation connections provide differential bridge-excitation voltage.
• Remote sensing continuously and automatically corrects for lead-wire induced
excitation voltage loss when using the RS connections.
• Shunt calibration can be used to correct for lead-wire induced desensitization of the
bridge.
sbRIO-9218 Getting Started Guide
|
© National Instruments Corporation
|
7