• Half-Bridge, requires the NI-9986
• Quarter-Bridge, requires the NI-9984 (120 Ω) or NI-9985 (350 Ω)
±16 V Connections
EX+
VEX+
VEX-
V
SUP+
V
SUP
-
sbRIO-9218
+
-
AI+
AI
-
/EX
-
The sbRIO-9218 provides optional 12 V sensor excitation. To use the 12 V excitation,
connect a 9 V DC to 30 V DC power supply to Vsup, connect the excitation terminals on
your sensor to EX+/EX-, and enable 12 V excitation in your software.