© National Instruments
|
A-19
RMX Programmable Power Supplies User Manual
Interface
Common specifications
Software protocol ..................................... IEEE Std 488.2-1992
Command language .................................. Complies with SCPI Specification 1999.0 Has a
compatibility mode (switchable)
1
.
Genesys series made by TDK-Lambda.
N5700 and N8700 made by Agilent
Technologies.
DSC series made by Sorensen.
PAG series made by Kikusui.
RS232C
Hardware .................................................. Complies with the EIA232D specifications
(excluding the connector)
RJ-45 connector
2
Baud rate: 1200, 2400, 4800, 9600, 19200,
38400, 57600, and 115200 bps
Data length: 8 bits, Stop bits: 1 bit,
Parity bit: None
No flow control
Program message terminator .................... LF during reception, CR/LF during transmission
USB
Hardware .................................................. Complies with the USB 2.0 specifications.
Baud rate: 480 Mbps (high speed).
Socket B type
Program message terminator .................... LF or EOM during reception, LF + EOM during
transmission
Device class .............................................. Complies with the USBTMC-USB488 device
class specifications
LAN
Hardware..............................................................IEEE 802.3 100Base-TX or 10Base-T Ethernet IPv4,
RJ-45 connector
Communication protocol .......................... VXI-11, HiSLIP, or SCPI-RAW
Program message terminator .................... VXI-11 and HiSLIP: LF or END during
reception, LF + END during transmission
SCPI-RAW: LF during reception, LF during
transmission
1
This setting does not guarantee compatibility with all measuring instrument application
software and drivers.
2
Category 5; use a straight cable.