Shock and Vibration
Operating shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC-60068-2-27. Meets MIL-
PRF-28800F Class 2 limits.)
Random vibration
Operating
5 Hz to 500 Hz, 0.3 g
rms
Nonoperating
5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-
PRF-28800F, Class 3.)
Safety
This product is designed to meet the requirements of the following standards of safety for
information technology equipment:
•
IEC 61010-1, EN 61010-1
•
CAN/CSA-C22.2 No. 61010-1
Note
For UL and other safety certifications, refer to the product label or the
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
AS/NZS CISPR 22: Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended
for use only in heavy-industrial locations.
Note
Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
8
|
ni.com
|
PXIe-8238 User Manual