background image

over time. NI uses these limits to reduce the probability that the instrument will be outside the
published specification limits at the end of the calibration cycle. Use the as-left limits when
performing verification after adjustment.

Verification

The following performance verification procedures describe the sequence of operation and
provide test points required to verify the NI PXIe-7865/7866. The verification procedures
assume that adequate traceable uncertainties are available for the calibration references.

Analog Input Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI
PXIe-7865/7866.
1.

Connect the NI R Series device to the calibrator as shown in the following figure.

Figure 1. Analog Input Verification Connections

Calibrator

HI

LO

AI1

+

AI0

+

2.

If your calibrator has a guard connection, connect that terminal to AI GND.

3.

Set the calibrator voltage to a Test Point value indicated in the appropriate 

Analog Input

Accuracy Verification Test Limits

 table in the 

Test Limits

 section of this document.

4.

Measure the analog input voltage on each channel by calling

nirio78xxCalibrationReadAIVoltage.vi

.

5.

Compare the measured voltage with the 1-year limits from the appropriate 

Analog Input

Accuracy Verification Test Limits

 table to verify the input accuracy. If the value is within

these limits, the device passes the test.

6.

Repeat steps 3 through 5 for each test point in the following 

Analog Input Accuracy

Verification Test Limits

 table.

4

  | 

ni.com

  | 

NI PXIe-7865 and NI PXIe-7866 Calibration Procedure

Содержание PXIe-7865

Страница 1: ...Device Drivers January 2020 or later Documentation Consult the following documents for information about your device LabVIEW and driver software All documents are available on ni com manuals and help...

Страница 2: ...ppm or less DMM NI 4070 Use a multiranging 6 digit DMM with an accuracy of 250 ppm Counter Agilent 53131A Use a counter accurate to 0 01 Low thermal copper EMF plug in cables Fluke 5440A 7002 Do not...

Страница 3: ...ew vi lib LabVIEW Targets FPGA RIO R Series 78XXR 783XR Calibration Public Self Calibration Complete the following steps to self calibrate the device Note No signal connections are needed for self cal...

Страница 4: ...to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections Calibrator HI LO AI1 AI0 2 If your calibrator has a guard connection connect that terminal to AI GND 3...

Страница 5: ...0 Negative full scale 5 5 000820 4 999180 5 001170 4 998830 2 2 Positive full scale 2 1 999661 2 000339 1 999521 2 000479 0 0 0 000137 0 000137 0 000137 0 000137 Negative full scale 2 2 000339 1 99966...

Страница 6: ...erification Test Limits table 6 Disconnect the DMM from the NI R Series device 7 Repeat steps 1 through 6 for all AO channels on the device Table 3 Analog Output Accuracy Verification Test Limits Rang...

Страница 7: ...ies Counter Verification Test Limits Set Point MHz Lower Limit MHz Upper Limit MHz 5 0 4 999500 5 000500 5 Stop the generation by calling nirio78xxCalibrationStopCounter vi 6 Disconnect the counter fr...

Страница 8: ...mines a new Requested Calibrator Value b Program the calibrator to output the Requested Calibrator Value At this point you must wait to ensure that the calibrator output has settled completely c Repea...

Страница 9: ...ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importan...

Страница 10: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...

Отзывы: