Test Conditions

The following setup and environmental conditions are required to ensure the device meets
published specifications.

Keep connections to the device as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.

Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ± 5 °C. The device temperature will be greater
than the ambient temperature.

Keep relative humidity below 80% .

Allow a warm-up time of at least 15 minutes to ensure that the device measurement
circuitry is at a stable operating temperature.

Initial Setup

Refer to your getting started guide for your device for information about how to install the
software and hardware.

Note

 

You can access all of the NI R Series VIs for calibration at 

<labview>

\vi.lib\LabVIEW Targets\FPGA\RIO\R Series\78XXR\783XR
\Calibration\Public

.

Self-Calibration

Complete the following steps to self-calibrate the device.

Note

 

No signal connections are needed for self-calibration.

1.

Wait 15 minutes for the device to warm-up.

2.

Launch the self-calibration VI 

nirio78xxCalibrationSelfCal.vi

.

3.

Call self-calibration before doing the first verification.

4.

Pass a TRUE to 

Save Cal Results

 to save the constants to the user portion of the

flash memory. If you pass False, the constants are then discarded.

As-Found and As-Left Limits

The as-found (1-Year) limits are the published specifications for the NI PXIe-7865/7866. NI
uses these limits to determine whether the NI PXIe-7865/7866 meets the specifications when it
is received for calibration. Use the as-found limits during initial verification.

The as-left (24-Hour) calibration limits are equal to the published NI specifications for the NI
PXIe-7865/7866, less guard bands for measurement uncertainty, temperature drift, and drift

NI PXIe-7865 and NI PXIe-7866 Calibration Procedure 

  | 

© National Instruments

  | 

3

Содержание PXIe-7865

Страница 1: ...Device Drivers January 2020 or later Documentation Consult the following documents for information about your device LabVIEW and driver software All documents are available on ni com manuals and help...

Страница 2: ...ppm or less DMM NI 4070 Use a multiranging 6 digit DMM with an accuracy of 250 ppm Counter Agilent 53131A Use a counter accurate to 0 01 Low thermal copper EMF plug in cables Fluke 5440A 7002 Do not...

Страница 3: ...ew vi lib LabVIEW Targets FPGA RIO R Series 78XXR 783XR Calibration Public Self Calibration Complete the following steps to self calibrate the device Note No signal connections are needed for self cal...

Страница 4: ...to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections Calibrator HI LO AI1 AI0 2 If your calibrator has a guard connection connect that terminal to AI GND 3...

Страница 5: ...0 Negative full scale 5 5 000820 4 999180 5 001170 4 998830 2 2 Positive full scale 2 1 999661 2 000339 1 999521 2 000479 0 0 0 000137 0 000137 0 000137 0 000137 Negative full scale 2 2 000339 1 99966...

Страница 6: ...erification Test Limits table 6 Disconnect the DMM from the NI R Series device 7 Repeat steps 1 through 6 for all AO channels on the device Table 3 Analog Output Accuracy Verification Test Limits Rang...

Страница 7: ...ies Counter Verification Test Limits Set Point MHz Lower Limit MHz Upper Limit MHz 5 0 4 999500 5 000500 5 Stop the generation by calling nirio78xxCalibrationStopCounter vi 6 Disconnect the counter fr...

Страница 8: ...mines a new Requested Calibrator Value b Program the calibrator to output the Requested Calibrator Value At this point you must wait to ensure that the calibrator output has settled completely c Repea...

Страница 9: ...ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importan...

Страница 10: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...