Note
Ensure that any connected external hardware is in an idle
state and not outputting signals during self-calibration. It is not
recommended to perform self-calibration while connected in direct
loopback configurations.
Note
For the best image rejection and residual LO performance, run Self
Cal or Self Calibrate Range for the image rejection and residual LO portion
any time the module temperature changes beyond ±1 °C.
Performing
PXIe-5842
Self-Calibration
You can self-calibrate your VST either interactively or programmatically; each
method yields the same self-calibration results.
•
Interactive methods
—Self-calibration in the InstrumentStudio application.
•
Programmatic methods
—Self-calibration in a supporting programming
environment using an API directly.
■
NI-RFSA
■
Using the RFSA Self-Calibration programming example
■
Using self-calibration functions
■
NI-RFSG: using self-calibration functions
■
RFmx: using self-calibration functions
To get started, choose one of the self-calibration methods, and self-calibrate the
PXIe-5842.
Performing a VST Self-Calibration Using
InstrumentStudio
NI recommends you perform an initial self-calibration using InstrumentStudio.