Performing a Self-Calibration Using NI-RFSA
NI recommends you perform the self-calibration using the NI-RFSA example when you self-
calibrate the module.
1.
Select
Start
»
All Programs
»
National Instruments
»
LabVIEW
to launch LabVIEW.
2.
Launch the Example Finder and navigate to the example.
a)
Select
Hardware Input and Output
»
Modular Instruments
»
NI-RFSA
»
Utilities
»
RFSA Self Calibration
.
b)
Open the example.
3.
Complete the following steps to configure the example:
a)
Select the device identifier assigned to the PXIe-3621 in MAX in the
[Resource
Name]
drop-down menu.
b)
Set Clock Source to
OnboardClock
.
c)
Set Self Calibration Step Operations to
Perform All Self Calibration Steps
.
4.
Run the VI.
12
|
ni.com
|
PXIe-5830 Getting Started Guide