NI PXIe-5611 Calibration Procedure
12
ni.com
•
Resolution bandwidth: 500 Hz
•
Sweep time: 5 ms
5.
Use the spectrum analyzer to measure the mean tone power at the
2nd harmonic frequency using the following spectrum analyzer
parameter settings:
•
Center Frequency: Frequency from step 3
×
2
•
Reference level: +5 dBm
•
Frequency span: 0 Hz
•
Resolution bandwidth: 500 Hz
•
Sweep time: 5 ms
6.
Calculate the power level of the harmonic in dBc by subtracting the
measured fundamental power from step 4 from the measured harmonic
power from step 5.
7.
Compare the calculated power level from step 6 to the corresponding
limit in Table 1.
8.
Repeat steps 3 through 7 for every frequency from 85 MHz to 2.2 GHz
in 5 MHz increments, including endpoints.
If the results are within the selected test limit, the device has passed this
portion of the verification.
Verifying LO Path Gain
Complete the following steps to verify the LO path gain of the NI 5611
module using a power meter.
1.
Connect the NI 5611 LO OUT front panel connector to the power
meter.
2.
Generate a signal with the following NI-RFSG property settings:
•
Frequency (Hz): 85 MHz
•
Power Level (dBm): 0 dBm
•
Generation Mode: CW
•
LO OUT Enabled: Enabled
3.
Use the power meter to measure the LO OUT power.
Table 3.
LO Filter Response Verification Test Parameters
Harmonic
2nd Harmonic Upper Limits (dBc)
F
≤
100 MHz
100 MHz < F
≤
2.2 GHz
2nd (2F)
–21
–31