NI PXIe-5611 Calibration Procedure
Note
This array is later used to compensate measurements and correct for splitter loss.
If you do not find the exact frequency needed in the splitter correction array, use the loss
corresponding to the next closest frequency.
You have successfully characterized your splitter.
Verification
This section describes the steps you must follow to verify successful
adjustment of the NI 5611.
Verification tests the following NI 5611 specifications:
•
LO filter response
•
LO path gain
•
Modulation impairments
•
Modulation bandwidth and impairments
•
RF path gain
•
Intermodulation products
Verification of the NI 5611 is complete only after you have successfully
completed all tests in this section.
Verifying LO Filter Response
Complete the following steps to verify the LO filter response of an NI 5611
module using a spectrum analyzer.
1.
Connect the NI 5611 LO OUT front panel connector to the spectrum
analyzer RF INPUT front panel connector.
2.
Connect the PXI Express chassis 10 MHz REF IN rear panel connector
to the spectrum analyzer REF OUT rear panel connector.
3.
Generate a signal with the following NI-RFSG property settings:
•
Frequency (Hz): 85 MHz
•
Power Level (dBm): 0 dBm
•
Generation Mode: CW
•
LO OUT Enabled: Enabled
4.
Use the spectrum analyzer to measure the mean fundamental tone
power using the following spectrum analyzer parameter settings:
•
Center Frequency: Frequency from step 3
•
Reference level: +5 dBm
•
Frequency span: 0 Hz