Table 1. PXIe-5413 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Precision
frequency
reference
PXIe-6674T or
Rubidium
reference
Verifications:
•
Timebase
accuracy
Adjustments:
•
Timebase
Frequency accuracy:
100 ppb
The ability to PLL lock to
the oscilloscope.
DMM
PXI-4071
Verifications:
•
DC voltage
accuracy
•
AC amplitude
accuracy
Adjustments:
•
DC voltage
DC voltage range: ±5 V
DC voltage accuracy:
<40 ppm
AC voltage accuracy:
±0.1% of reading at 50 kHz
AC Input Range: 0.05 V
RMS
to 5 V
RMS
AC Input Impedance:
≥10 MΩ
Bandwidth: ≥100 kHz
Double banana
plug-to-BNC (f)
Pasternak PE9008
Verifications:
•
DC voltage
accuracy
•
AC amplitude
accuracy
Adjustments:
•
DC voltage
Impedance: 50 Ω
SMA (m)-to-
BNC (m) cable
—
Verifications:
•
DC voltage
accuracy
•
AC amplitude
accuracy
Adjustments:
•
DC voltage
Frequency range: DC to
1 MHz
Impedance: 50 Ω
Length: <1 m