Table 4. DC Accuracy Verification (Continued)
Config
Input
Impedance
(
Ω
)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Sample
Rate
(GS/s)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits (mV)
30
50
1 V
4.5
3.0
4.975
±37.7
±17.8275
31
50
1 V
-4.5
1.5
-4.975
±37.7
±17.3525
32
50
1 V
-4.5
3.0
-4.975
±37.7
±17.8275
33
50
2 V
0
1.5
0.95
±39.2
±19.17
34
50
2 V
0
3.0
0.95
±39.2
±19.835
35
50
2 V
0
1.5
-0.95
±39.2
±19.17
36
50
2 V
0
3.0
-0.95
±39.2
±19.835
37
50
2 V
4
1.5
4.95
±55.2
±26.37
38
50
2 V
4
3.0
4.95
±55.2
±27.035
39
50
2 V
-4
1.5
-4.95
±55.2
±26.37
40
50
2 V
-4
3.0
-4.95
±55.2
±27.035
41
50
4 V
0
1.5
1.9
±78.2
±38.72
42
50
4 V
0
3.0
1.9
±78.2
±38.15
43
50
4 V
0
1.5
-1.9
±78.2
±38.72
44
50
4 V
0
3.0
-1.9
±78.2
±38.15
45
50
4 V
3
1.5
4.9
±90.2
±44.12
46
50
4 V
3
3.0
4.9
±90.2
±43.55
47
50
4 V
-3
1.5
-4.9
±90.2
±44.12
48
50
4 V
-3
3.0
-4.9
±90.2
±43.55
49
50
10 V
0
1.5
4.75
±195.2
±100.125
50
50
10 V
0
3.0
4.75
±195.2
±95.375
51
50
10 V
0
1.5
-4.75
±195.2
±100.125
52
50
10 V
0
3.0
-4.75
±195.2
±95.375
PXIe-5113 Calibration Procedure
|
© National Instruments
|
13