NI PXIe-4480/4481 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures for the PXIe-4480. Table 2 lists the same for the PXIe-4481. If the recommended
equipment is not available, select a substitute using the minimum requirements listed in the
table.
Table 1.
Recommended Equipment (PXIe-4480)
Equipment
Recommended
Model(s)
Where Used
Minimum
Requirements
Calibrator
Fluke 5700A
AI Gain
Frequency Range:
40 Hz to 200 kHz
Voltage Range:
up to 6.3 V
rms
ACV Accuracy:
40 Hz: 0.017%
1 kHz: 0.016%
20 kHz: 0.017%
50 kHz: 0.037%
100 kHz: 0.086%
200 kHz: 0.32%
Function Generator
Keysight (Agilent)
33250A series
Timebase
Frequency
Frequency Range:
up to 90 kHz
Frequency Accuracy:
2 ppm
Voltage Range:
up to 9 V
pk
Digital Multimeter
PXI-4070
IEPE Current
Current Accuracy:
0.9%
InfiniBand 12X-6BNC
Cable
National
Instruments
SHB12X-6BNC
All
N/A
BNC Shorting Cap
(quantity 6)
Pomona
Electronics 5085
or
Pomona
Electronics
3840-50
AI Offset
Resistance: ≤50 Ω