NI PXIe-4480/4481 Calibration Procedure
|
© National Instruments
|
5
Table 2.
Recommended Equipment (PXIe-4481)
Equipment
Recommended
Model(s)
Where Used
Minimum
Requirements
Calibrator
Fluke 5700A
AI Gain
Frequency Range:
40 Hz to 200 kHz
Voltage Range:
up to 6.3 V
rms
ACV Accuracy:
40 Hz: 0.017%
1 kHz: 0.016%
20 kHz: 0.017%
50 kHz: 0.037%
100 kHz: 0.086%
200 kHz: 0.32%
Function Generator
Keysight (Agilent)
33250A series
Timebase
Frequency
Frequency Range:
up to 90 kHz
Frequency Accuracy:
2 ppm
Voltage Range:
up to 9 V
pk
InfiniBand 12X-6BNC
Cable
National
Instruments
SHB12X-6BNC
All
N/A
BNC Shorting Cap
(quantity 6)
Pomona
Electronics 5085
or
Pomona
Electronics
3840-50
AI Offset
Resistance: ≤50 Ω
BNC (F) to Banana Plug
Adapter
Pomona
Electronics 1269
AI Gain
N/A