background image

NI PXIe-4339/TB-4339 Installation Guide and Terminal Block Specifications

|

© National Instruments

|

17

Storage Environment

Ambient temperature range .............................. -40 °C to 71 °C 

(Tested in accordance with IEC 60068-2-1 and 
IEC 60068-2-2. Meets MIL-PRF-28800F 
Class 3 limits.)

Relative humidity range.................................... 5% to 95%, noncondensing 

(Tested in accordance with IEC 60068-2-56.)

Shock and Vibration

Operating shock................................................ 30 g peak, half-sine, 11 ms pulse 

(Tested in accordance with IEC 60068-2-27. 
Meets MIL-PRF-28800F Class 2 limits.)

Random vibration

Operating .................................................. 5 Hz to 500 Hz, 0.3 g

rms

Non-operating........................................... 5 Hz to 500 Hz, 2.4 g

rms

 

(Tested in accordance with IEC 60068-2-64. 
Non-operating test profile exceeds the 
requirements of MIL-PRF-28800F, Class 3.)

Safety

Measurement Category..................................... I

1

Caution

Do 

not

 connect the TB-4339/B/C to signals or use for measurements 

within Measurement Categories II, III or IV.

Caution

The protection provided by the TB-4339/B/C can be impaired if it is used 

in a manner not described in this document.

Safety Standards

This product meets the requirements of the following standards of safety for electrical equipment 
for measurement, control, and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1, CSA 61010-1

Note

For UL and other safety certifications, refer to the product label or the 

Online 

Product Certification

 section.

1   

Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not 
intended for direct connection to the MAINS building installations of Measurement Categories CAT II, 
CAT III, or CAT IV.

Содержание PXIe-4339

Страница 1: ...nnecting to other modules that could be damaged by the voltage present on the terminal block However you should only use these terminal blocks with their supported modules Contents Electromagnetic Com...

Страница 2: ...ns when the product is connected to a peripheral device or test object or if the product is used in residential or commercial areas To minimize interference with radio and television reception and pre...

Страница 3: ...tallation Install the Software Note You must be an administrator to install NI software and devices on your computer Ensure that the following software is installed before installing the SC Express ha...

Страница 4: ...rted PXI Express slot in the chassis SC Express devices can be placed only in PXI Express Peripheral slots PXI Express Hybrid Peripheral slots and PXI Express System Timing slots Refer to the chassis...

Страница 5: ...4 while completing the following steps Note You can find the pinout names and locations in Measurement Automation Explorer MAX at any time by right clicking the device name under Devices and Interfac...

Страница 6: ...to determine the function of the terminal The NI PXIe 4339 User Manual provides more detailed wiring information Make sure no exposed wire extends past the screw terminal Exposed wire increases the r...

Страница 7: ...raps to connect the shielded signal wires to the tie wrap holes for additional strain relief when necessary 10 Reinstall the top cover and tighten the captive top cover screws Note For information abo...

Страница 8: ...the bottom of the terminal block 3 Tighten the mounting screw at the top of the TB 4339 B C to attach it to the NI PXIe 4339 module Note For safety purposes and to prevent damage to equipment when hi...

Страница 9: ...annels and tasks and getting started developing an application For detailed NI software version support refer to the NI DAQmx Readme 1 Launch MAX 2 Expand Devices and Interfaces and expand the chassis...

Страница 10: ...ction for instructions for creating NI DAQmx simulated devices 3 Right click the module name and select Self Test When the self test completes a verification message appears If an error occurs refer t...

Страница 11: ...the DAQ Assistant in MAX Complete the following steps to create a task using the DAQ Assistant in MAX 1 In MAX right click Data Neighborhood and select Create New to open the DAQ Assistant 2 Select N...

Страница 12: ...ional examples refer to zone ni com Terminal Block Removal To remove the terminal block refer to Figure 7 while completing the following steps Note The chassis can be powered on or off when removing t...

Страница 13: ...so can damage the module To remove the module refer to Figure 8 while completing the following steps Note To maintain forced air cooling in the PXIe system refer to the Maintain Forced Air Cooling No...

Страница 14: ...c protective bag Figure 8 Removing SC Express Modules Create a Simulated Device To run examples without the hardware installed use an NI DAQmx simulated device To create a simulated device in MAX 1 La...

Страница 15: ...If you need to return your National Instruments hardware for repair or device calibration go to ni com info and enter rdsenn to start the Return Merchandise Authorization RMA process Specifications T...

Страница 16: ...free from contaminants before returning it to service Environmental Specifications Maximum altitude 2 000 m 800 mbar at 25 C ambient temperature Pollution Degree 2 Indoor use only Operating Environmen...

Страница 17: ...e requirements of MIL PRF 28800F Class 3 Safety Measurement Category I1 Caution Do not connect the TB 4339 B C to signals or use for measurements within Measurement Categories II III or IV Caution The...

Страница 18: ...aland per CISPR 11 Class A equipment is intended for use only in heavy industrial locations Note Group 1 equipment per CISPR 11 is any industrial scientific or medical equipment that does not intentio...

Страница 19: ...nts website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assi...

Страница 20: ...s You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal expor...

Отзывы: