PXIe-4310 Calibration Procedure
|
© National Instruments
|
9
5.
Save the adjustment constants to the EEPROM, using the DAQmx Close External
Calibration function with the following parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_Commit
This function also saves the date, time, and temperature of the adjustment to the onboard
memory.
Note
If an error occurs during adjustment, no constants will be written to the
EEPROM.
6.
Place the calibrator in Standby.
7.
Disconnect the calibrator from the device.
The device is now adjusted with respect to your external source.
Reverification
Repeat the
section using the
in Table 4 to determine the
as-left status of the device.
Note
If any test fails Reverification after performing an adjustment, verify that you
have met the
before returning your device to NI. Refer to
for assistance in returning the device to NI.
Specifications
Refer to the
PXIe-4310 Specifications
document for detailed PXIe-4310 specification information.