9

5.

Save the adjustment constants to the EEPROM, using the DAQmx Close External 
Calibration function with the following parameters:

calHandle

calHandle

action

DAQmx_Val_Action_Commit

This function also saves the date, time, and temperature of the adjustment to the onboard 
memory.

Note

If an error occurs during adjustment, no constants will be written to the 

EEPROM.

6.

Place the calibrator in Standby.

7.

Disconnect the calibrator from the device.

The device is now adjusted with respect to your external source.

Reverification

Repeat the 

Accuracy Verification

 section using the 

24-Hour Limits

 in Table 4 to determine the 

as-left status of the device.

Note

If any test fails Reverification after performing an adjustment, verify that you 

have met the 

Test Conditions

 before returning your device to NI. Refer to 

Worldwide 

Support and Services

 for assistance in returning the device to NI.

Specifications

Refer to the 

PXIe-4310 Specifications 

document for detailed PXIe-4310 specification information.

Worldwide Support and Services

The NI website is your complete resource for technical support. At 

ni.com/support

, you 

have access to everything from troubleshooting and application development self-help resources 
to email and phone assistance from NI Application Engineers.

Visit 

ni.com/services

 for information about the services NI offers.

Visit 

ni.com/register

 to register your NI product. Product registration facilitates technical 

support and ensures that you receive important information updates from NI.

NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 
78759-3504, USA. For up-to-date contact information for your location, visit 

ni.com/

contact

.

Содержание PXIe-4310

Страница 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Страница 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Страница 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Страница 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Страница 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Страница 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Страница 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Страница 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Страница 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Страница 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: