PXIe-4310 Calibration Procedure

|

© National Instruments

|

5

Test Conditions

The following setup and environmental conditions are required to ensure the PXIe-4310 meets 
published specifications.

Review and become familiar with the entire procedure before beginning the calibration 
process.

Keep connections to the PXIe-4310 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-4310 (10V) are secure.

Use shielded copper wire for all cable connections to the TB-4310 (10V). Use twisted-pair 
wire to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C. The PXIe-4310 temperature will be 
greater than the ambient temperature.

Keep relative humidity below 80%.

For valid test limits, maintain the device temperature within ±1 °C from the last 
self-calibration and ±10 °C from the last external calibration. The device temperature will 
be greater than the ambient temperature. Refer to the 

Checking Device Temperature 

Changes

 section for more information about verification and adjustment temperatures and 

temperature drift.

Allow a warm-up time of at least 15 minutes to ensure that the PXIe-4310 measurement 
circuitry is at a stable operating temperature.

Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters are 
clean, and that the empty slots contain filler panels. For more information, refer to the 

Maintain Forced-Air Cooling Note to Users

 document available at 

ni.com/manuals

.

Initial Setup

Refer to the 

PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Getting Started Guide and 

Terminal Block Specifications

 for information about how to install the software and hardware 

and how to configure the device in NI Measurement & Automation Explorer (MAX).

Note

When a device is configured in MAX, it is assigned a device identifier. Each 

function call uses this identifier to determine which DAQ device to verify. This 
document uses 

Dev1

 to refer to the device name. In the following procedures, use the 

device name as it appears in MAX.

Self-Calibration

A self-calibration must be performed before performing device verification. Perform a 
self-calibration after the device has warmed up for the recommended time period of 15 minutes.

You can initiate self-calibration using MAX, by completing the following steps.
1.

Launch MAX.

2.

Select 

My System»Devices and Interfaces»PXIe-4310

.

Содержание PXIe-4310

Страница 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Страница 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Страница 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Страница 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Страница 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Страница 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Страница 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Страница 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Страница 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Страница 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: