PXIe-4302/4303 and TB-4302C Calibration Procedure
4.
Refer to Table 4 to configure the PXI-4071 and acquire a voltage measurement.
5.
Acquire a voltage measurement with the PXIe-4302/4303.
a.
Create a DAQmx task.
b.
Create and configure the AI channel according to the values shown in Table 5.
Table 3.
PXIe-4139 Voltage Output Setup
Configuration
Value
Function
Voltage output
Sense
Remote
Range
600 mV range for test points less than 100 mV
60 V range for all other test points
Current Limit
20 mA
Current Limit Range
200 mA
Table 4.
PXI-4071 Voltage Measurement Setup
Configuration
Value
Function
DC measurement
Range
1 V range for test points less than 100 mV.
10 V range for all other test points.
Digital Resolution
7.5 digits
Aperture Time
100 ms
Autozero
On
ADC Calibration
On
Input Impedance
> 10 G
DC Noise Rejection
High Order
Number of Averages
1
Power Line Frequency
Dependent upon the local power line characteristics.