PXIe-4302/4303 and TB-4302C Calibration Procedure

|

© National Instruments Corporation

|

5

Test Conditions

The following setup and environmental conditions are required to ensure the PXIe-4302/4303 
meets published specifications.

Keep connections to the PXIe-4302/4303 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-4302 are secure.

Use shielded copper wire for all cable connections to the TB-4302. Use twisted-pair wire 
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C. The PXIe-4302/4303 temperature will be 
greater than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the PXIe-4302/4303 
measurement circuitry is at a stable operating temperature.

Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters are 
clean, and that the empty slots contain filler panels. For more information, refer to the 

Maintain Forced-Air Cooling Note to Users

 document available on 

ni.com

.

Initial Setup

Refer to the 

NI PXIe-4302/4303 and TB-4302/4302C Installation Guide and Terminal Block 

Specifications

 for information about how to install the software and hardware and how to 

configure the device in Measurement & Automation Explorer (MAX).

Note

When a device is configured in MAX, it is assigned a device identifier. Each 

function call uses this identifier to determine which DAQ device to verify or to verify 
and adjust. This document uses 

Dev1

 to refer to the device name. In the following 

procedures, use the device name as it appears in MAX.

Accuracy Verification

The following performance verification procedures describe the sequence of operations and 
provides the test points required to verify the PXIe-4302/4303. The verification procedures 
assume that adequate traceable uncertainties are available for the calibration references.

The PXIe-4302/4303 has 32 independent analog input channels. The input range of each channel 
can be set to 10 V or 100 mV. You can verify the accuracy of either range for all of the channels 
depending upon your desired test coverage.

Complete the following steps to verify the voltage mode accuracy of the PXIe-4302/4303.
1.

Set the PXIe-4139 voltage output to zero volts.

2.

Connect the PXIe-4139 and PXI-4071 to the TB-4302 as shown in Figure 2.

3.

Use Table 3 to configure the PXIe-4139 to output a Test Point value for the appropriate 
range shown in Table 6, beginning with the values in the first row.

Содержание PXIe-4302

Страница 1: ...date 9 Reverification 10 TB 4302C Verification 10 Test Equipment 10 Accuracy Verification 11 Specifications 14 Revision History 14 NI Services 14 Software Calibrating the PXIe 4302 4303 requires you t...

Страница 2: ...ting the PXIe 4302 4303 Test Equipment Table 1 lists the equipment recommended for the performance verification and adjustment procedures of the PXIe 4302 4303 If the recommended equipment is not avai...

Страница 3: ...SMU PXIe 4139 Noise 0 1 Hz to 10 Hz peak to peak is 60 V or better at 10 V Noise 0 1 Hz to 10 Hz peak to peak is 2 V or better at 100 mV Table 1 Recommended Equipment for PXIe 4302 4303 Verification a...

Страница 4: ...th 1 or better tolerance to build a voltage divider to bias the PXIe 4139 output and set the common mode input of the PXIe 4302 4303 to zero volts Connect one resistor between AI and AIGND and the oth...

Страница 5: ...n Guide and Terminal Block Specifications for information about how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Note When a device is co...

Страница 6: ...Voltage output Sense Remote Range 600 mV range for test points less than 100 mV 60 V range for all other test points Current Limit 20 mA Current Limit Range 200 mA Table 4 PXI 4071 Voltage Measurement...

Страница 7: ...onfiguration Value Channel Name Dev1 ai0 32 Task AI voltage Sample Mode Finite samples Sample Clock Rate 5000 Samples per Channel 5000 Maximum Value Appropriate maximum range value from Table 6 Minimu...

Страница 8: ...with the following parameters Device In Dev1 Password NI1 4 Call the 4302 4303 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output f...

Страница 9: ...eps 5 through 8 for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the...

Страница 10: ...returning the device to NI TB 4302C Verification This section provides information for verifying the performance of the TB 4302C Test Equipment Table 8 lists the equipment recommended for verifying t...

Страница 11: ...VSUP VSUP A9 AI0 VSUP VSUP AI19 VSUP AI18 VSUP AI20 AI21 AI23 VSUP AI22 VSUP VSUP VSUP AI16 AI17 VSUP VSUP AI28 VSUP AI27 VSUP AI29 AI30 AI31 VSUP AI24 VSUP VSUP VSUP A25 AI26 VSUP VSUP S N 158141A 0...

Страница 12: ...annel to Shunt Reference Designator Correlation Channel Shunt Reference Designator CH0 R10 CH1 R11 CH2 R12 CH3 R13 CH4 R14 CH5 R15 CH6 R16 CH7 R17 CH8 R21 CH9 R20 CH10 R19 CH11 R18 CH12 R25 CH13 R24 C...

Страница 13: ...1 6 Compare the results to the Lower Limit and Upper Limit values in Table 11 If the results are between these values the device passes the test 7 Repeat steps 3 through 6 for all other 5 shunt resist...

Страница 14: ...Customers The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014...

Отзывы: