Ensure the switches' certifications meet your test application requirements. NI
recommends UL-certified safety switches to ensure reliability.

Install the switches in accordance with the switch manufacturer specifications.

Test the switches periodically to ensure proper implementation and reliability.

Safety Guidelines for System Operation

Caution

 

Hazardous voltage of up to the maximum voltage of the device may

appear at the output terminals if the safety interlock terminal is closed. Open the
safety interlock terminal when the output connections are accessible. With the safety
interlock terminal open the output voltage level/limit is limited to ±40 VDC, and
protection will be triggered if the voltage measured between the device HI and LO
terminals exceeds ±(42 Vpk ±0.4 V).

Caution

 

Do not apply voltage to the safety interlock connector inputs. The

interlock connector is designed to accept passive normally open contact closure
connections only.

To ensure a system containing the NI 4136/4137 is safe for operators, components, or
conductors, take the following safety precautions:

Ensure proper warnings and signage exists for workers in the area of operation.

Provide training to all system operators so that they understand the potential hazards and
how to protect themselves.

Inspect connectors, cables, switches, and any test probes for any wear or cracking before
each use.

Before touching any of the connections to the high terminal or high sense on the
NI 4136/4137, discharge all components connected to the measurement path. Verify with
a DMM before interaction with connections.

Installing the NI 4136/4137

Caution

 

To prevent damage to the device caused by ESD or contamination, handle

the device using the edges or the metal bracket.

1.

Ensure the AC power source is connected to the chassis before installing the modules.

The AC power cord grounds the chassis and protects it from electrical damage while you
install the modules.

2.

Power off the chassis.

3.

Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.

4.

Remove the black plastic connectors from all the captive screws on the module front
panel.

5.

Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.

NI PXIe-4136/4137 Getting Started Guide

  | 

© National Instruments

  | 

7

Содержание PXIe-4136

Страница 1: ...ay If the product is damaged return it to NI for repair Contents Electromagnetic Compatibility Guidelines 2 Safety Guidelines for Hazardous Voltages 2 Verifying the System Requirements 2 Unpacking the...

Страница 2: ...aution To ensure the specified EMC performance the length of all I O cables must be no longer than 3 m 10 ft Safety Guidelines for Hazardous Voltages If hazardous voltages are connected to the device...

Страница 3: ...Never touch the exposed pins of connectors Note Do not install a device if it appears damaged in any way 3 Unpack any other items and documentation from the kit Store the device in the antistatic pack...

Страница 4: ...ring the Environment Ensure that the environment you are using the NI 4136 4137 in meets the following specifications Operating Environment Ambient temperature range 0 C to 55 C Tested in accordance w...

Страница 5: ...s you to restart shut down or restart later System Safety Guidelines Safety Guidelines for System Design and Implementation The NI 4136 4137 is capable of generating hazardous voltages and working wit...

Страница 6: ...entry points to the test fixture enclosure enabling hazardous voltage ranges on the instrument Figure 2 System Level Connection Typical 1 2 Safety Interlock Connector Test Fixture Enclosure Operator...

Страница 7: ...re proper warnings and signage exists for workers in the area of operation Provide training to all system operators so that they understand the potential hazards and how to protect themselves Inspect...

Страница 8: ...e module guides at the top and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 4 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Han...

Страница 9: ...ing Screw Connected to Drain Wire 3 Region Where Exposed Wiring Is Allowed 7 62 mm 300 in 4 Region Free from Exposed Wiring Cable Ground Shield or Drain Wire 8 89 mm 350 in Minimum 5 Output Connector...

Страница 10: ...In the configuration tree expand Devices and Interfaces to see the list of installed devices Installed devices appear under the name of their associated chassis 3 Expand your Chassis tree item MAX li...

Страница 11: ...thorization RMA Testing with the NI DCPower Soft Front Panel 1 Disconnect the output connector from the NI 4136 4137 front panel 2 Ensure that the safety interlock input on the test fixture is closed...

Страница 12: ...wer configures and operates the device hardware and performs basic acquisition and measurement options using LabVIEW VIs or LabWindows CVI functions C or LabWindows CVI Available at Program Files IVI...

Страница 13: ...that the NI 4136 4137 appears in MAX If the ACCESS LED fails to illuminate after you power on the PXI Express chassis a problem may exist with the PXI Express power rails a hardware module or the LED...

Страница 14: ...nity ni com community Located using the NI Example Finder NI PXIe 4136 Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support...

Страница 15: ...3504 National Instruments also has offices located around the world For telephone support in the United States create your service request at ni com support or dial 1 866 ASK MYNI 275 6964 For telepho...

Страница 16: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Отзывы: