13. Get the spectrum information using the niRFSA Get Spectral Info for SMT VI or the
niRFSA_GetSpectralInfoForSMT
function.
14. Measure power in band using the SMT Power In Band VI or the
SmtSpectrumUnitConversion
function with the following parameter settings:
•
units
to
dBm
•
impedance
(called
impedanceOhm
in the function) to
50
•
center frequency
to the equivalent RF frequency for the spur from
Non-Input-Related Spurs Test Frequencies
•
bandwidth
to
500
This value is the
NI 5663 Spurious Level
.
15. Compare the
NI 5663 Spurious Level
to the verification test limits in Table 12.
16. Repeat steps 9 to 15 for all settings in Table 11.
17. Close the NI 5663 session using the niRFSA Close VI or the
niRFSA_close
function.
If the results are within the selected test limit, the device has passed this portion of the
verification.
Sideband Spurs Verification
Table 16 shows the test limits.
1.
Connect the RF source to the power splitter input through the 3.5 mm-to-3.5 mm cable.
2.
Connect the power splitter reference output to power sensor A.
3.
Connect the other power splitter output to the NI 5601 RF IN connector through the 3.5 mm
(m) to 3.5 mm (m) adapter. Figure 7 shows the completed equipment setup.
Table 12.
Non-Input-Related Spurs Verification Test Limits
Frequency
Test Limit
10 MHz to 6.6 GHz
<-95 dBm