NI PXI-5650/5651/5652 Calibration Procedure
|
© National Instruments
|
27
Adjustment
Following the adjustment procedure automatically updates the calibration date and temperature
in the EEPROM of the PXI-5650/5651/5652.
Adjustment involves the following PXI-5650/5651/5652 procedures:
•
Automatic level control (ALC)
•
Attenuators
•
ALC limits
•
Frequency accuracy
Note
Adjustment procedures
must
be performed in the order listed.
Adjusting ALC
This procedure measures the response of the ALC circuit. This response is used to generate
accurate power levels at the PXI-5650/5651/5652 RF OUT front panel connector. Complete the
following steps to adjust the ALC of the PXI-5650/5651/5652 using the power meter thermal
sensor:
1.
Connect the PXI-5650/5651/5652 RF OUT front panel connector to the power meter
thermal sensor.
2.
Call the niRFSG Initialize External Calibration VI.
3.
Call the niRFSG 5652 Update ALC Limits Table VI, and set the
update max power spec
parameter to TRUE and the
update other table limits
parameter to FALSE.
4.
Call the niRFSG 5652 Initialize ALC Calibration VI, and set the
measurement range
parameter to
Above 10 MHz
.
5.
Call the niRFSG 5652 Configure ALC Calibration VI.
6.
Use the power meter to measure the RF OUT power at the frequency specified by the
frequency to measure
parameter of the niRFSG 5652 Configure ALC Calibration VI.
7.
Call the niRFSG 5652 Adjust ALC Calibration VI, and wire the measured value from step 6
to the
measured RF OUT power
parameter.
8.
Repeat steps 5 through 7 until the
ALC calibration complete
parameter of the
niRFSG 5652 Adjust ALC Calibration VI returns a value of TRUE.
9.
Connect the PXI-5650/5651/5652 RF OUT front panel connector to the power meter
thermal sensor.
10. Call the niRFSG 5652 Initialize ALC Calibration VI, and set the
measurement range
parameter to
Below 10 MHz
.
11. Call the niRFSG 5652 Configure ALC Calibration VI.
12. Use the power meter to measure the RF OUT power at the frequency specified by the
frequency to measure
parameter of the niRFSG 5652 Configure ALC Calibration VI.
13. Call the niRFSG 5652 Adjust ALC Calibration VI, and wire the measured value from
step 12 to the
measured RF OUT power
parameter.