10
|
ni.com
|
NI PXI-5650/5651/5652 Calibration Procedure
4.
Use the power meter to measure the RF OUT power. Compare the measured RF OUT
power to the
Test Limit
listed in Table 4.
Table 4.
High Power Level Accuracy Test Points and Limits
Device
Frequency
Start Power
(dBm)
Test Limit (dB)
PXI-5650/5651/
5652
500 kHz
5
±1.6
1 MHz
5
2 MHz
5
5 MHz
5
10 MHz
8
±0.75
20 MHz
8
50 MHz
10
100 MHz
10
150 MHz
10
200 MHz
10
500 MHz
10
1 GHz
10
PXI-5651/5652
1.5 GHz
10
±0.75
2 GHz
8
2.5 GHz
8
3 GHz
8
3.3 GHz
7
PXI-5652
3.5 GHz
7
±1.0
4 GHz
5
4.5 GHz
5
5 GHz
0
5.5 GHz
0
6 GHz
0
6.6 GHz
0