12. Power on the chassis.

Related Information

Installing the Software

 on page 6

Chassis Slot Configurations

The following figures show the two possible slot configurations for the NI 5672 module in an
8-slot NI PXIe-1062Q hybrid chassis.

Figure 4.

 NI 5672 Module in Slots 2, 3, and 4 of an NI PXIe-1062Q Chassis

2

6

3

3

4

5

4

8

1

7

NI PXIe-1062Q

1. System Controller Expansion Slots
2. PXI Express System Controller Slot
3. PXI Peripheral Slots (PXI only)
4. PXI Express Hybrid Peripheral Slots

5. PXI Express System Timing Slot (PXI Express

only)

6. NI PXI-5610 Module
7. NI PXIe-5442 Module
8. PXI/PXI Express Filler Panels

8

  | 

ni.com

  | 

NI PXIe-5672 Getting Started Guide

Содержание PXI-5610

Страница 1: ...rograms National Instruments NI RFSG Documentation The specifications document for your device is installed with the driver software Refer to ni com manuals for the most recent specifications for your device Caution The protection provided by this product may be impaired if it is used in a manner not described in this document Hot Surface If the NI 5672 has been in use it may exceed safe handling ...

Страница 2: ...cial areas To minimize interference with radio and television reception and prevent unacceptable performance degradation install and use this product in strict accordance with the instructions in the product documentation Furthermore any modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Caution To e...

Страница 3: ...on Never touch the exposed pins of connectors Notify NI if the device appears damaged in any way Do not install a damaged device 3 Unpack any other items and documentation from the kit Store the device in the antistatic package when the device is not in use NI PXIe 5672 Getting Started Guide National Instruments 3 ...

Страница 4: ...G Module 3 Two Semi Flexible SMA m to SMB Cables 4 Two Cable Wrenches Part Number 746016 01 5 Screwdriver Part Number 772006 01 6 Driver Software DVD 7 NI PXIe 5672 Getting Started Guide this document 8 Read Me First Safety and Electromagnetic Compatibility 9 Maintain Forced Air Cooling Note to Users 4 ni com NI PXIe 5672 Getting Started Guide ...

Страница 5: ... 5672 in meets the following specifications Operating ambient temperature NI PXI 5610 upconverter 0 C to 55 C Tested in accordance with IEC 60068 2 1 and IEC 60068 2 2 Meets MIL PRF 28800F Class 3 low temperature limit and MIL PRF 28800F Class 2 high temperature limit NI PXIe 5442 AWG 0 C to 55 C operating in all National Instruments PXI Express chassis Operating relative humidity IEC 60068 2 56 1...

Страница 6: ... NI RFSG before installing the hardware Before you install the hardware refer to the guidelines in the Maintain Forced Air Cooling Note to Users included with the module to ensure that the chassis can cool the device effectively To use the included cables you must install the NI 5442 AWG module in the slot immediately to the right of the NI 5610 RF upconverter module 1 Ensure the AC power source i...

Страница 7: ...lace the module edges into the module guides at the top and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 3 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatched Position 9 Latch the module in place by pulling up on the ejector handle 10 Secure the device front panel to the chassis using the front panel mounti...

Страница 8: ...d chassis Figure 4 NI 5672 Module in Slots 2 3 and 4 of an NI PXIe 1062Q Chassis 2 6 3 3 4 5 4 8 1 7 NI PXIe 1062Q 1 System Controller Expansion Slots 2 PXI Express System Controller Slot 3 PXI Peripheral Slots PXI only 4 PXI Express Hybrid Peripheral Slots 5 PXI Express System Timing Slot PXI Express only 6 NI PXI 5610 Module 7 NI PXIe 5442 Module 8 PXI PXI Express Filler Panels 8 ni com NI PXIe ...

Страница 9: ...Slot 3 PXI Peripheral Slots PXI only 4 PXI Express Hybrid Peripheral Slots 5 PXI Express System Timing Slot PXI Express only 6 NI PXI 5610 Module 7 NI PXIe 5442 Module 8 PXI PXI Express Filler Panels Note The NI 5610 module blocks the PXI Express System Timing slot in the configuration shown in the preceding figure NI PXIe 5672 Getting Started Guide National Instruments 9 ...

Страница 10: ...Note Ensure that the SMA cable center pins properly align with the SMA connectors to avoid bending the cable pins or damaging the front panel connectors on the device 2 Using a flexible SMA to SMB cable connect the TO AWG CLK IN connector on the NI 5610 front panel to the CLK IN connector on the NI 5442 front panel 3 Hand tighten all SMA cable ends on the SMA connectors The cable connectors should...

Страница 11: ...ll modules If an LED is not illuminated refer to the Troubleshooting section of this document Related Information Configuring the NI 5672 in MAX on page 16 Direct Connections to the NI 5672 The NI 5672 is a precision RF instrument that is sensitive to ESD and transients Ensure you take the following precautions when making direct connections to the NI 5672 to avoid damaging the device Note Do not ...

Страница 12: ... two multicolor LEDs Figure 7 NI 5610 RF Upconverter Module Front Panel Table 1 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance 12 ni com NI PXIe 5672 Getting Started Guide ...

Страница 13: ...her The output signal at these connectors is always on and cannot be disabled 10 MHz OUT REF IN Routes an external frequency reference signal that can be propagated to the PXI backplane The NI 5610 can lock to this signal INPUT Connects to the CH 0 output on the NI 5442 AWG module front panel Routes the IF signal from the NI 5442 AWG module for frequency translation LOCAL OSC OUT 0 Routes the auxi...

Страница 14: ...the device GREEN The module is ready to be programmed by NI RFSG ACTIVE Indicates the status of the module PLLs OFF The module is in an uninitialized state there is no error AMBER The module PLLs are attempting to lock GREEN The module is in a ready state applicable PLLs are locked and the reverse power protection circuit is closed RED The module has detected an error state this state may indicate...

Страница 15: ...ulticolor LEDs Figure 8 NI 5442 AWG Module Front Panel Table 4 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance NI PXIe 5672 Getting Started Guide National Instruments 15 ...

Страница 16: ...a problem with a power rail AMBER The module is being accessed Accessed means that the device setup registers are being written to in order to control the device or load waveforms GREEN The module is ready to be programmed by NI RFSG ACTIVE Indicates the state of the module OFF The module is not armed or triggered AMBER The module is armed and waiting for a Start Trigger GREEN The module has recei...

Страница 17: ...amming the NI 5672 5 Associate the hardware modules that comprise your device a Select the NI 5610 that is identified as not configured in the configuration tree b In the Associated Devices section select the appropriate module from each system component drop down listbox For the NI 5672 you must associate the NI 5610 RF upconverter module with the NI 5442 AWG module c Click Save in the MAX toolba...

Страница 18: ...es of the NI 5672 in MAX NI RFSG operates the modules as a single instrument LabWindows CVI Available at Program Files IVI Foundation IVI Drivers niRFSG Microsoft Visual C C Add all required include and library files to your project to create an NI RFSG application in Microsoft Visual C C Related Information For detailed instructions about how to generate signals in a specific ADE refer to the Get...

Страница 19: ...shooting section of this document if an ACTIVE LED does not turn on or if the NI RFSG SFP generates an error During signal generation the ACTIVE LED on each of the NI 5672 hardware modules illuminates 5 Click RF On Off to stop signal generation Building a Basic NI RFSG Application You can build a basic NI RFSG application in LabVIEW for generating continuous sine wave signals Adding the Core NI RF...

Страница 20: ...Panel 11 In the VI front panel power level dBm control enter 0 In the frequency Hz control enter 100M 100 MHz 12 In the VI front panel resource name control enter the NI 5610 upconverter device name that you specified in MAX Adding a While Loop Add a While Loop to continuously generate the signal and check the generation status until you click the Stop button 1 Display the VI block diagram and sel...

Страница 21: ...of the status element to an input of the Or function 9 Right click the unused input of the Or function and select Create Control to create a Boolean control 10 In the VI front panel right click the Boolean control created in step 9 and select Replace Modern Boolean Stop Button to create a Stop button Figure 12 While Loop with Stop Button Adding an Error Indicator Add an error indicator to the VI f...

Страница 22: ...verter module name specified in MAX in the resource name control 4 Click the Run button on the toolbar to initiate sine wave generation 5 Click the VI front panel STOP button to stop sine wave generation You have successfully generated a continuous sine wave signal using the NI RFSG instrument driver and the NI 5672 22 ni com NI PXIe 5672 Getting Started Guide ...

Страница 23: ...einstall a damaged device 4 Reinstall the module in a different chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform a self test What Should I Do if the NI 5672 Doesn t Appear in MAX 1 In the MAX configuration tree click Devices and Interfaces 2 Expand the Chassis tree to see the list of installed devices and press F5 to refresh the list...

Страница 24: ...d module in a different slot 5 Power on the chassis 6 Perform the self test again What Should I Do if the Thermal Shutdown Error Appears The thermal shutdown error appears when device temperatures exceed safe limits The NI 5672 shuts down until temperatures fall to acceptable levels and you reset the device in MAX 1 Power off the chassis that contains the device 2 Review the Maintain Forced Air Co...

Страница 25: ...re EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example Finder Tip The NI RF Signal Generators Help is an HTML version of a traditional user manual that includes detailed information about RF fundamentals device features and programming with NI RFSG Worldwi...

Страница 26: ...t information support phone numbers email addresses and current events Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for information on National Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the appropriate location Help ...