2
|
ni.com
|
NI PXI-4130 Calibration Procedure
To calibrate the NI 4130, you must install the NI-DCPower instrument driver version 1.2 or later
on the Windows calibration system. You can download the latest version of NI-DCPower at
ni.com/downloads
. NI-DCPower supports LabVIEW and LabWindows
™
/CVI
™
. When you
install NI-DCPower, you need to install only the application software that you intend to use.
LabVIEW support is in the
nidcpower.llb file
, and all calibration VIs are accessible from
the NI-DCPower Calibration palette. For LabWindows/CVI users, the
nidcpower.fp
provides access to the calibration functions. Refer to Table 1 for file locations.
C function calls are valid for any compiler capable of calling a 32-bit or 64-bit DLL. Many of
the functions use constants defined in the
nidcpower.h
file. To use these constants in C, you
must include
nidcpower.h
in the calibration program.
For more information about calibration VIs and functions, refer to the
Programming with
NI-DCPower
book in the
NI DC Power Supplies and SMUs Help
. The
NI DC Power Supplies
and SMUs Help is
accessible at
Start»All Programs»National Instruments»NI-DCPower»
Documentation»NI DC Power Supplies and SMUs Help
.
Table 1.
Calibration File Locations (NI-DCPower 1.2 or Later)
File Name and Location
Description
IVI\Bin\nidcpower_32.dll
NI-DCPower driver containing the entire
NI-DCPower API, including calibration functions.
IVI\Bin\nidcpower_64.dll
NI-DCPower driver containing the entire
NI-DCPower API, including calibration functions.
IVI\Lib\msc\nidcpower.lib
NI-DCPower library for Microsoft C containing
the entire NI-DCPower API, including calibration
functions.
<LabVIEW
>\instr.lib\
niDCPower\niDCPower.llb
LabVIEW VI library containing VIs for calling the
NI-DCPower calibration API. You can access
calibration functions from the NI-DCPower
calibration section of the LabVIEW function
palette.
IVI\Drivers\niDCPower\
nidcpower.fp
LabWindows/CVI function panel file that includes
calibration function prototypes and help on using
NI-DCPower in the LabWindows/CVI
environment.
IVI\Include\nidcpower.h
NI-DCPower header file, which you must include
in any C program accessing calibration functions.
This file includes the entire NI-DCPower API,
including calibration functions.