For Your Safety

Caution

Before undertaking any troubleshooting, maintenance, or exploratory procedure, read carefully the 

WARNING and CAUTION notices.

This equipment contains voltage hazardous to human life and safety, and is capable of inflicting personal injury.

Mainframe Grounding

—The PXI-1002 requires a connection from the premise wire safety ground to the PXI-1002 

chassis ground. The earth safety ground must be connected during use of this equipment to minimize shock hazards. 

Live Circuits

—Operating personnel and service personnel must not remove protective covers when operating or 

servicing the PXI-1002.

Explosive Atmosphere

—Do 

not

 operate the mainframe in conditions where flammable gases are present. Under such 

conditions this equipment is unsafe and may ignite the gases or gas fumes.

Modification

—Do 

not 

modify any part of the mainframe from its original condition. Unsuitable modifications may 

result in safety hazards.

Connecting Safety Ground

Warning

The PXI-1002 chassis is designed with a three-position NEMA 15-5 style plug that connects the 

ground line to the chassis ground. To minimize shock hazard, make sure your electrical power outlet has an 
appropriate earth safety ground that is connected whenever you power up the chassis.

If your power outlet does 

not

 have an appropriate ground connection, you must connect the premise wire safety ground to 

the chassis grounding screw located on the rear panel. To connect the safety ground, complete the following steps:

1.

Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw on the rear of the chassis using a toothed grounding 
lug. The wire must have green insulation with a yellow stripe or must be noninsulated (bare).

2.

Attach the opposite end of the wire to permanent earth ground using toothed washers or a toothed lug.

Specifications

AC Input Specifications for AC-Only 
Power Supply

Input voltage range.........90–135/180–265 VAC

auto select

Input frequency range ...47 to 63 Hz

Input current......................8 A @ 115 VAC; 5 A @ 230 VAC

Power disconnect.............The power switch causes the power 

module to supply DC power to the 
PXI-1002 backplane. The power cord 
provides main power disconnect.

Note

There are no user-replaceable fuses.

Safety

UL 3111-1, IEC 1010-1, CSA 22.2 No. 1010.1; Installation 
Category II; Pollution Degree 2; Safety Class 1

DC Output Specifications for AC-Only 

Power Supply

Maximum power.............................. 170 W

DC Current Capacity (IMP)

Voltage

IMP (Steady-State Current)

+3.3 V*

14 A

+5 V*

20 A

+12 V

4 A

–12 V

0.8 A

*Total combined power from +3.3 V and +5 V is 112 W max.

Environmental

For indoor use only.
Operating temperature ................. 0 to 50 °C ambient
Storage temperature ...................... –40 to +70 °C
Operating relative humidity ........ 10 to 90%, noncondensing

Backplane

Backplane bare-board material: UL 94 V-0 recognized (File No. E 96467); Backplane connectors: Conform to IEC 917 and IEC 
1076-4-101, and are UL 94 V-0 rated

!