NI TB-9214 Calibration Procedure

|

© National Instruments

|

7

Figure 3.  

NI TB-9214 Verification Connections

3.

Reinstall the top cover on the NI TB-9214.

4.

Connect the NI TB-9214 to the NI 9214.

5.

Power on the thermocouple calibrator and set the thermocouple type to E-type.

6.

Wait 15 minutes or the warm-up time required by the thermocouple calibrator, whichever 
is longer.

7.

Acquire a CJC

temp0

 value.

a.

Create and configure an AI Temp Built-In Sensor task for CJC channel _cjtemp0 on 
the NI 9214 according to Table 2.

b.

Configure the AI Temp Built-In Sensor task for CJC channel _cjtemp0 according to 
Table 3.

Table 2.  

NI 9214 CJC Channel Configuration

Physical Channels

Timing Mode

_cjtemp0

High Resolution

Table 3.  

NI 9214 CJC Timing Configuration

Sample Mode

Samples Per Channel

Rate (S/s)

Finite

5

1

.

.

.

NI TB-9214

B

a

rrier

S

trip

E0+

E0–

E1+

E14–

E15+

E15–

E1–

E14+

.

.

.

.

.

.

TC0+

TC0–

+

+

TC1+

TC14–

TC15+

TC15–

TC1–

TC14+

.

.

.

Mini TC
Connector

TC
C

a

li

b

r

a

tor

R

a

pid Temper

a

t

u

re Tr

a

n

s

ient

I

s

ol

a

tion Enclo

su

re

Содержание NI TB-9214

Страница 1: ...ts Software 1 Documentation 2 Test Equipment 3 Building the Barrier Strip Assembly 4 Test Conditions 5 Initial Setup 6 Verification 6 Terminal Block CJC Verification 6 Where to Go for Support 9 Softwa...

Страница 2: ...ons and Specifications NI TB 9214 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Страница 3: ...Module Use an NI 9214 module that is verified to be within product specifications Chassis NI cDAQ 9178 BarrierStrip Assembly Thermocouple Wire TT E 24 SLE 100 E type SLE 24 AWG 20 to 30 meters Thermo...

Страница 4: ...Crimp the chromel spade lugs to the chromel wire to create 16 positive spade lugs Crimp the constantan spade lugs to the constantan wire to create 16 negative spade lugs 2 Use the thermocouple wire sp...

Страница 5: ...I TB 9214 meets published specifications Keep connections to the NI TB 9214 as short as possible Long cables and wires act as antennae picking up extra noise that can affect measurements Verify that a...

Страница 6: ...er 6 Connect the power source to the NI cDAQ 9178 chassis 7 Launch Measurement Automation Explorer MAX 8 Right click the device name and select Self Test to ensure that the module is working properly...

Страница 7: ...temp0 value a Create and configure an AI Temp Built In Sensor task for CJC channel _cjtemp0 on the NI 9214 according to Table 2 b Configure the AI Temp Built In Sensor task for CJC channel _cjtemp0 ac...

Страница 8: ...fore acquiring data Wait at least 300 ms for the Ectron 1140A to settle before acquiring data with the NI 9214 Disregard the first six seconds of the acquisition to compensate for the open thermocoupl...

Страница 9: ...at the NI TB 9214 is outside of the limits refer to Where to Go for Support for assistance in returning the device to NI Where to Go for Support The National Instruments Web site is your complete reso...

Отзывы: