background image

NI 1357/1358/1359 Kit Installation Guide

8

ni.com

3.

Select 

NI-DAQmx SCXI Chassis

, and choose the chassis. 

The Create New SCXI Chassis dialog box opens as shown in Figure 7.

Figure 7.  

Configuring the Create New SCXI Dialog Box

4.

Create a new SCXI chassis:

Note

If you are using an NI PXI-1010 chassis and the switch controller is 

not

 in the 

rightmost PXI slot, create the chassis as an NI SCXI-1000 in MAX.

a.

Select the switch controller (NI USB-1359 or 
NI 4021/4060/407

x

) cabled to the communicating SCXI switch 

module from the 

Chassis Communicator 

listbox. If MAX 

detects only one communicating device, this device is selected as 
the chassis communicator by default and the listbox is dimmed.

b.

SCXI kits only—Select the module slot connected to the chassis 
communicator from the 

Communicating SCXI Module Slot

 

listbox.

c.

Enter the chassis address setting in the 

Chassis Address 

field. 

Ensure that the setting you enter matches the address setting on 
the SCXI chassis as shown in Figure 6.

d.

Enable the auto-detect option to auto-detect installed SCXI switch 
modules. If you do not auto-detect switch modules, MAX disables 
the Communicating SCXI Module Slot listbox.

e.

Click 

Save

.

After creating a new chassis, the SCXI Chassis Configuration dialog 
box opens as shown in Figure 8. The 

Modules 

tab is selected by 

default.

Содержание NI SCXI 1357

Страница 1: ...nal Instruments USB 1359 Manual Get Pricing Availability at ApexWaves com Call Today 1 800 915 6216 Email sales apexwaves com https www apexwaves com modular systems national instruments accessories USB 1359 ...

Страница 2: ...ls among switches Introduction Note Both SCXI and USB kits connect to an SCXI switch with a 10 pin rear connector Refer to the NI Switches Getting Started Guide to determine the switches and switch controllers that your kit supports The NI 1359 kit 2 slots consists of the following components NI USB 1359 2 slot high voltage analog bus HVAB backplane adapter with USB connector or NI SCXI 1359 2 slo...

Страница 3: ...NI DAQ 7 4 or later before you install the NI 1357 1358 1359 kit Complete the following steps to install the NI 1357 1358 1359 kit 1 Power off and unplug the SCXI chassis Follow proper electrostatic discharge ESD precautions and ensure high voltage analog backplane is not connected before installing the hardware 1 8 Slot HVAB Backplane Adapter 16 Screws 2 2 Slot HVAB Backplane Adapter NI USB 1359 ...

Страница 4: ... the switch controller Refer to Figures 2 and 3 to verify the connector positions on the NI 1359 Figure 2 2 Slot HVAB Backplane Adapter Connectors Figure 3 8 Slot HVAB Backplane Adapter Connectors 1 HVAB Connectors 2 9 Pin DIN Connectors 3 USB Connector 4 10 Pin Rear Connector 1 HVAB Connectors 2 9 Pin DIN Connectors 3 10 Pin Rear Connector 300V MAX CAT II HV ANALOG BUS N114 300V MAX CAT II HV ANA...

Страница 5: ...its Connect one end of the SH9MD AUX cable to the NI SCXI 1359 cable adapter and the other end to the NI 4021 4060 407x switch controller as shown in Figure 4 Figure 4 Cabling an NI 4021 4060 407x Switch Controller to a 10 Pin Rear Connector Switch Module 1 SCXI Switch 2 HVAB Connector 3 10 Pin Rear Connector 4 SH9MD AUX Cable 5 NI 4021 4060 407x 6 NI SCXI 1359 7 NI SCXI 1000 1001 or NI PXI 1010 C...

Страница 6: ... configure the switch module Multichassis Configuration Complete the following steps to create a multichassis switch configuration 1 Complete the Installing the Kit section steps 1 through 7 for each chassis you want to add to the configuration 2 Set a distinct address for each chassis Figure 6 illustrates the address switches located on the front of the chassis 1 SCXI Switch 2 HVAB Connector 3 10...

Страница 7: ...N of the HVAB backplane adapter and the other end to the NI 4021 4060 407x switch controller b Using another SH9MD AUX cable connect the AUX OUT of the HVAB backplane adapter on the first chassis to the AUX IN of the HVAB backplane adapter on the next chassis c Repeat step b until the HVAB backplane adapters on each chassis are connected USB kits For every NI USB 1359 in the configuration connect ...

Страница 8: ...gure SCXI switch modules cabled to an NI SCXI 1357 1358 1359 kit proceed directly to step 4 to create a new SCXI chassis If you are using NI DAQmx to configure SCXI switch modules cabled to an NI USB 1357 1358 1359 kit and the NI USB 1359 is auto detected and the chassis is auto created in MAX expand the NI DAQmx Devices tab right click the chassis and choose Properties The SCXI Chassis Configurat...

Страница 9: ...f MAX detects only one communicating device this device is selected as the chassis communicator by default and the listbox is dimmed b SCXI kits only Select the module slot connected to the chassis communicator from the Communicating SCXI Module Slot listbox c Enter the chassis address setting in the Chassis Address field Ensure that the setting you enter matches the address setting on the SCXI ch...

Страница 10: ...dit the default name click the Device Identifier textbox and enter a unique alphanumeric ID for the switch module c If you are using a connected accessory such as a terminal block select it from the Accessory listbox 6 Click Details to configure the switch module options The Details dialog box for the switch module configuration dialog box opens The Terminal Block Topology tab is selected by defau...

Страница 11: ...switch module information A message at the top of the SCXI Chassis Configuration dialog box displays the configuration status If a warning appears in the message box you can save the configuration If an error appears in the message box you cannot save the configuration 11 To test the configuration right click the chassis name listed under Devices and Interfaces and select Test After the test a mes...

Страница 12: ... NI SCXI 1129 150 V CAT I Example 2 NI SCXI 1000 chassis NI SCXI 1359 2 slot backplane adapter behind Slots 3 and 4 NI SCXI 1127 300 V CAT II in Slot 4 NI SCXI 1129 150 V CAT I in Slot 2 The NI SCXI 1127 does not share the HVAB with the NI SCXI 1129 Therefore the NI SCXI 1127 can operate at the full rating 300 V CAT II Example 3 NI SCXI 1000 chassis NI SCXI 1359 2 slot backplane adapter behind Slo...

Страница 13: ...0 1 UL 61010 1 CAN CSA C22 2 No 61010 1 Note For UL and other safety certifications refer to the product label or visit ni com certification search by model number or product line and click the appropriate link in the Certification column Electromagnetic Compatibility Emissions EN 55011 Class A at 10 m FCC Part 15A above 1 GHz Immunity EN 61326 1997 A2 2001 Table 1 CE C Tick and FCC Part 15 Class ...