PXIe-5673E Calibration Procedure
To calibrate the PXIe-5673E, you must install NI-RFSG version 1.8 or later on the calibration
system. You can download the latest version of NI-RFSG at
ni.com/downloads
.
NI-RFSG supports programming the calibration procedures in LabVIEW. Refer to Table 1 for
file locations.
Documentation
You might find the following documentation helpful as you perform the calibration procedure:
•
PXIe-5673E Specifications
•
PXIe-5673E Getting Started Guide
•
NI RF Signal Generators Help
, including LabVIEW programming references
The latest versions of these documents are available on
ni.com/manuals
.
This calibration procedure calibrates the PXIe-5673E as a single device. To calibrate the
PXIe-5450/5451 Waveform Generator, PXIe-5611 IQ Modulator, or PXIe-5650/5651/5652
RF Analog Signal Generator individually, refer to their calibration procedures at
ni.com/
manuals
.
Calibration Interval
The measurement accuracy requirements of your application determine how often you
should calibrate your device. NI recommends that you perform a complete calibration for
the PXIe-5673E at least once a year. You can shorten this calibration interval based on the
accuracy demands of your application.
Table 1.
Calibration File Locations
File Name and Location
Description
<IVI>
\Bin\niRFSG.dll
NI-RFSG driver containing the entire NI-RFSG API,
including calibration functions.
<LabVIEW>
\instr.lib\
niRFSG\niRFSG.llb
LabVIEW VI library containing VIs for calling the
NI-RFSG calibration API. You can access calibration VIs
from the NI-RFSG Calibration section of the LabVIEW
Functions palette.