Table 8.
NI 5652 RF Signal Generator Module Front Panel Connectors
Connector
Use
RF OUT
Output terminal for the RF signal at the requested frequency and power level.
REF IN/OUT Routes a frequency reference signal to or from the NI 5652 module.
This connector can be configured as an external frequency reference input or
as an output for the onboard 10 MHz Reference Clock.
Table 9.
NI 5652 RF Signal Generator Module Front Panel LEDs
LED
Indications
ACCESS Indicates the basic hardware status of the NI 5652 module.
OFF—The module is not yet functional or has detected a problem with a PXI
power rail.
AMBER—The module is being accessed.
Accessed
means that the device setup
registers are being written to in order to control the device.
GREEN—The module is ready to be programmed.
ACTIVE Indicates the state of the NI 5652 module.
OFF—The module is not generating a signal.
AMBER—The module PLLs are attempting to lock.
GREEN—The module is generating a signal; applicable PLLs are locked.
RED—The module has detected an error state; this may indicate a PLL lock
failure or a thermal shutdown condition.
Configuring the NI 5663 in MAX
Use Measurement & Automation Explorer (MAX) to configure your National Instruments
hardware. MAX informs other programs about which devices reside in the system and how
they are configured. MAX is automatically installed with NI-RFSA.
1. Launch MAX.
2. In the configuration tree, double-click
Devices and Interfaces
to see the list of installed
devices.
Note
If you are using the NI 5663 with the LabVIEW Real-Time Module,
expand
Remote Systems
. Find your target IP address or name, expand it, and
then expand
Devices and Interfaces
.
16
|
ni.com
|
NI PXIe-5663 Getting Started Guide