manualshive.com logo in svg
background image

10

|

ni.com

|

NI PXIe-5653 Calibration Procedure

Verifying LO1 Amplitude Accuracy

This verification ensures that the internal LO1 circuitry is adjusted for correct amplitude 
accuracy. Complete the following procedure to determine the as-found status of the NI 5653.
1.

Connect the NI 5653 LO1 front panel connector to the power meter.

2.

Create a new device session for the NI 5653.

3.

Create a list of test frequencies from 3.2 GHz to 8.3 GHz in 100 MHz increments, including 
endpoints.

4.

Calculate the expected output power, 

P

Expected

, for each frequency using the following 

formulas.
For all frequencies from 3.2 GHz to 8.2 GHz:

For the frequency 8.3 GHz:

P

Expected

 = 6.5 dBm

5.

Set the RF frequency to the first value of the frequency array you created in step 3 and 
commit the settings to hardware.

6.

Check the signal generation status and verify that there are no reported errors or warnings. 

7.

After correcting for the measurement frequency, use the power meter to measure the 
NI 5653 LO1 front panel connector power.

8.

Calculate the power deviation, 

P

Deviation

, using the following formula:

9.

Repeat steps 5 through 8 for each point in the frequency array you created in step 3.

10. Verify that the power deviation results calculated in step 9 for each point are within ±2.5 dB 

for LO1.

11. Close the device session.

Verifying LO2 Amplitude Accuracy

This verification ensures that the internal LO2 circuitry is adjusted for correct amplitude 
accuracy. Complete the following procedure to determine the as-found status of the NI 5653.
1.

Connect the NI 5653 LO2 front panel connector to the power meter.

2.

Create a new device session for the NI 5653.

3.

Set the RF frequency to 4 GHz and commit the settings to hardware.

4.

Check the signal generation status and verify that there are no reported errors or warnings.

5.

After correcting for the measurement frequency, use the power meter to measure the LO2 
output power.

6.

Ensure that the output power is between 6.5 dBm and 13 dBm.

7.

Close the device session.

P

Expected

10.5 3

frequency GHz

(

)

3.2

GHz

5.0

GHz

---------------------------------------------------------------------

dBm

(

)

=

P

Deviation

dB

(

)

P

Measured

dBm

(

)

P

Expected

dBm

(

)

=

Содержание NI PXIe-5653

Страница 1: ...racy 7 Verifying 100 MHz Reference Amplitude Accuracy 8 Verifying LO1 Frequency Accuracy 9 Verifying LO1 Amplitude Accuracy 10 Verifying LO2 Amplitude Accuracy 10 Verifying LO3 Amplitude Accuracy 11 V...

Страница 2: ...ent environments ADE When you install the software you need to install support only for the ADE that you intend to use Documentation You might find the following documents helpful as you perform the c...

Страница 3: ...g LO3 Phase Noise Adjusting Reference Accuracy Adjusting YIG Frequency Accuracy Frequency range 9 MHz to 10 GHz Frequency accuracy 1 ppb over the frequency range 900 MHz to 9 GHz recommended Noise flo...

Страница 4: ...m Frequency range 9 MHz to 8 3 GHz Accuracy 0 5 Linearity Up to 15 dBm MA2475 only SMA m to SMA m Cable Verifying 10 MHz Reference Frequency Accuracy Verifying 100 MHz Reference Frequency Accuracy Ver...

Страница 5: ...and the calibrator into the same power strip to avoid ground loops Use a torque wrench appropriate for the type of RF connector that you are using NI recommends a 0 565 N m 5 lb in wrench for SMA conn...

Страница 6: ...ncy Accuracy This verification ensures that the internal oven controlled crystal oscillator OCXO is adjusted for correct frequency accuracy Complete the following procedure to determine the as found s...

Страница 7: ...the signal generation status and verify that there are no reported errors or warnings 6 After correcting for the measurement frequency use the power meter to measure the NI 5653 REF OUT 10 MHz front...

Страница 8: ...ng 100 MHz Reference Amplitude Accuracy This verification ensures that the 100 MHz reference circuitry is adjusted for correct amplitude accuracy Complete the following procedure to determine the as f...

Страница 9: ...and commit the settings to hardware 6 Check the signal generation status and verify that there are no reported errors or warnings 7 Configure the signal source analyzer as follows Center frequency Fre...

Страница 10: ...he power meter to measure the NI 5653 LO1 front panel connector power 8 Calculate the power deviation PDeviation using the following formula 9 Repeat steps 5 through 8 for each point in the frequency...

Страница 11: ...nternal LO1 circuitry is adjusted for correct phase noise accuracy Complete the following procedure to determine the as found status of the NI 5653 1 Connect the NI 5653 LO1 front panel connector to t...

Страница 12: ...Ensure that the recorded measurements are within the limits set in Table 3 Verifying LO2 Phase Noise This verification ensures that the internal LO2 circuitry is adjusted for correct phase noise accu...

Страница 13: ...verify that there are no reported errors or warnings 8 Use the signal source analyzer to measure the phase noise Record the results 9 Close the device session 10 Ensure that the recorded measurements...

Страница 14: ...zer to perform a manual frequency sweep at the settings listed in Table 6 5 Set the RF frequency to 4 GHz and commit the settings to hardware 6 If your signal source analyzer does not automatically ce...

Страница 15: ...This adjustment yields a more accurately tuned VCO frequency 1 Connect the NI 5653 REF OUT 10 MHz front panel connector to the signal source analyzer RF IN front panel connector 2 Connect the signal...

Страница 16: ...nce using a signal source analyzer 1 Connect the NI 5653 LO1 front panel connector to the signal source analyzer RF IN front panel connector 2 Connect the signal source analyzer REF IN rear panel conn...

Страница 17: ...ameter returns a value of TRUE 7 Close the external calibration session Store the results to the EEPROM on the device 8 Disconnect the NI 5653 LO1 front panel connector from the power meter and connec...

Страница 18: ...uracy This reverification ensures that the 100 MHz reference circuitry is adjusted for correct amplitude accuracy following adjustment Complete the following procedure to determine the as left status...

Страница 19: ...is reverification ensures that the internal LO2 circuitry is adjusted for correct amplitude accuracy following adjustment Complete the following procedure to determine the as left status of the NI 565...

Страница 20: ...t status of the device Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshoot...

Отзывы: