![National Instruments NI PXIe-5450 Скачать руководство пользователя страница 168](http://html2.mh-extra.com/html/national-instruments/ni-pxie-5450/ni-pxie-5450_user-manual_3486740168.webp)
Chapter 5
Signal Generation Fundamentals
NI PXIe-5450 User Manual
5-16
ni.com
Now, Analog Filter 2 can easily filter out all the images due to the digital
generation of the signal. This behavior is seen in the frequency domain
representation in the previous figure and in the time domain representation
in the following figure.
Note
The allowable range of interpolation factors is dependent on the NI signal generator
being used.
Using two times interpolation filtering with a DAC effective sample rate of
2f
s
eliminates images well and generates a good signal. However,
increasing the interpolation filter to 4 further improves the output signal.
Analog Filter 2
Analog Filter 1
f
s
0
2f
s
3
f
s
0.5f
s
Images
f
0
Signal
Power
4f
s
Without Interpolation
With Interpolation