NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
17
3.
Generate a waveform with NI PXIe-4463:
a.
Configure an AO voltage channel task using the values in Table 22.
b.
Call
Basic Function Generator
using the parameters in Table 23.
c.
Write the resulting waveform to AO, and start the task.
4.
Measure the signal with the frequency counter, and compare the results with the limits in
Table 24.
5.
Stop the task.
Table 22.
AO Setup
Configuration
Value
Physical Channels
Dev_name
/ao0
AO Gain
0 dB
Output Terminal Configuration
Differential
Rate
51200 S/s
Sample Mode
Continuous Sample
Samples Per Channel
500
Table 23.
Basic Function Generator Parameters
Parameter
Value
Signal Type
Sine Wave
Frequency
20480 Hz
Amplitude
10 V
Sampling Info»Sampling Rate
51200 S/s
Sampling Info»Number of Samples
500
Table 24.
Timebase Frequency Verification Test Limits
Programmed
Frequency (Hz)
As Found
Test Limit
As Left
Test Limit
Min (Hz)
Max (Hz)
Min (Hz)
Max (Hz)
20480
20479.556
20480.444
20479.662
20480.338