Verification
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.
Related Information
on page 18
Repeat the
Verification
section to determine the as-left status of the device.
Required Software
Calibrating the NI 4141 requires you to install the following software on the calibration
system:
•
NI-DCPower. The NI 4141 was first supported in NI-DCPower 1.7.5.
•
Supported application development environment (ADE)—LabVIEW or
LabWindows
™
/CVI
™
•
Supported operating system—Windows
You can download all required software from
Related Documentation
You might find the following documents helpful as you perform the calibration procedure:
•
NI DC Power Supplies and SMUs Getting Started Guide
•
NI DC Power Supplies and SMUs Help
•
NI PXIe-4141 Specifications
•
NI-DCPower Readme
Visit
for the latest versions of these documents.
Password
The default password for password-protected operations is
NI
.
Calibration Interval
............................................................................
Recommended calibration interval
1 year
Test Equipment
The following table lists the equipment NI recommends for the performance verification and
adjustment procedures. If the recommended equipment is not available, select a substitute
using the minimum requirements listed in the table.
2
|
ni.com
|
NI PXIe-4141 Calibration Procedure