National Instruments Corporation
23
NI PXI-5660 Getting Started Guide
Measurements
For more information about making measurements with the RF vector signal analyzer, refer to the
NI PXI-5660 Help
.
RF Vector Signal Analyzer Amplitude Reading Does Not Match Source
If the RF vector signal analyzer reports an incorrect amplitude reading, check for the following possible
errors:
1.
Check that the discrepancy between the RF vector signal analyzer and the source is within the error
limits of the devices, as follows:
a.
Verify the absolute amplitude accuracy of the RF vector signal analyzer.
b.
Verify the output level accuracy of the source.
2.
Check for loss in the cables, which can be substantial. For example, RG58 coaxial cable loses about
2.1 dB of signal amplitude per foot at 2 GHz. Unless you are using high-quality cables, expect
losses when working with high-frequency signals.
3.
Ensure that you are using the amplitude calibration functions in the NI Spectral Measurements
Toolkit. Failure to apply calibration functions can result in large amplitude errors.
4.
If errors continue, contact NI technical support or visit
ni.com/support
.
Technical Support and Professional Services
Visit the following sections of the award-winning National Instruments Web site at
ni.com
for
technical support and professional services:
•
Support
—Technical support at
ni.com/support
includes the following resources:
–
Self-Help Technical Resources
—For answers and solutions, visit
ni.com/support
for
software drivers and updates, a searchable KnowledgeBase, product manuals, step-by-step
troubleshooting wizards, thousands of example programs, tutorials, application notes,
instrument drivers, and so on. Registered users also receive access to the NI Discussion
Forums at
ni.com/forums
. NI Applications Engineers make sure every question submitted
online receives an answer.
–
Standard Service Program Membership
—This program entitles members to direct access
to NI Applications Engineers via phone and email for one-to-one technical support as well as
exclusive access to on demand training modules via the Services Resource Center. NI offers
complementary membership for a full year after purchase, after which you may renew to
continue your benefits.
For information about other technical support options in your area, visit
ni.com/services
,
or contact your local office at
ni.com/contact
.
•
Training and Certification
—Visit
ni.com/training
for self-paced training, eLearning virtual
classrooms, interactive CDs, and Certification program information. You also can register for
instructor-led, hands-on courses at locations around the world.
•
System Integration
—If you have time constraints, limited in-house technical resources, or other
project challenges, National Instruments Alliance Partner members can help. To learn more, call
your local NI office or visit
ni.com/alliance
.
•
Declaration of Conformity (DoC)
—A DoC is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system affords
the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the
DoC for your product by visiting
ni.com/certification
.