NI Switches Getting Started Guide
After you have prepared your chassis, complete the instructions appropriate
to your controller. Refer to Table 2 to determine if your controller supports
and connects to your SCXI switch module.
E Series
E Series devices can directly cable only to switches with 50-pin rear
connectors. Refer to Table 2 to determine if your switch module has a
50-pin rear connector.
1.
Power off and unplug the SCXI chassis. Follow proper ESD
precautions to ensure you are grounded before installing the hardware.
2.
Remove the filler panel in an unused SCXI slot.
3.
Touch any metal part of your chassis to discharge static electricity.
4.
Insert the device into the SCXI slot.
5.
Screw the device front panel to the chassis front panel mounting rail.
6.
Attach the SCXI-1349 cable adapter to the rear of the SCXI switch
module.
7.
Screw the cable adapter to the chassis cable adapter mounting rail.
8.
Connect one end of the SH6868 cable to the SCXI-1349 cable adapter
and the other to the E Series device.
Figure 6.
Cabling to an E Series Device
1
SCXI Switch
2
SCXI-1000/1001 or PXI-1010 Chassis
3
SCXI-1349 Cable Adapter
4
E Series Device
5
SH6868 Cable
1
2
5
4
3