NI 9242 User Manual and Specifications
|
© National Instruments
|
41
Input voltage range (AI
x
and Neutral-to-GND, AI
x
-to-Neutral)
Typical........................................ 500 V
pk
Minimum.................................... 497 V
Overvoltage withstand ...................... 500 V
rms
continuous,
600 V
rms
for 10 s
Surge withstand ................................ 5 kV (1.2 µs/50 µs)
Input coupling................................... DC
Input impedance, AI
x
-to-Ground
and Neutral-to-GND ......................... 1 M
Ω
Table 1.
DC and AC Accuracy
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error)
Calibrated max (-40 °C to 70 °C)
0.26%
0.14%
Calibrated typ (23 °C ±5 °C)
0.05%
0.022%
Uncalibrated max (-40 °C to 70 °C)
0.50%
0.26%
Uncalibrated typ (23 °C ±5 °C)
0.18%
0.06%
*
Range equals 354 V (250 V
rms
×
√
2)