Channel-to-earth ground
Continuous
60 VDC, Measurement
Category I
Withstand
1,000 V, verified by a 5 s
dielectric withstand test
Measurement Category I is for measurements performed on
circuits not directly connected to the electrical distribution system
referred to as
MAINS
voltage. MAINS is a hazardous live
electrical supply system that powers equipment. This category is
for measurements of voltages from specially protected secondary
circuits. Such voltage measurements include signal levels, special
equipment, limited-energy parts of equipment, circuits powered
by regulated low-voltage sources, and electronics.
Note
Measurement Categories CAT I and CAT O are
equivalent. These test and measurement circuits are not
intended for direct connection to the MAINS building
installations of Measurement Categories CAT II,
CAT III, or CAT IV.
6
|
ni.com
|
NI 9239 Getting Started Guide
Содержание NI 9239
Страница 1: ...NI 9239...
Страница 2: ...GETTING STARTED GUIDE NI 9239 4 AI 10 V 24 Bit 50 kS s ch Simultaneous...