Figure 1. NI-9238 Cable Connections for EMC Compliance
NI 9238
Shield
AI0+
AI0–
Ferrite
Shield
AI1+
AI1–
Ferrite
Shield
AI2+
AI2–
Ferrite
Signal
Source
Signal
Source
Signal
Source
Shield
AI3+
AI3–
Ferrite
Signal
Source
Related concepts
■
Special Conditions for Marine Applications
Some products are approved for marine (shipboard) applications. To verify marine
approval certification for a product, visit
, search by
model number, and click the appropriate link.
Notice
In order to meet the EMC requirements for marine applications,
install the product in a shielded enclosure with shielded and/or filtered
power and input/output ports. In addition, take precautions when
designing, selecting, and installing measurement probes and cables to
ensure that the desired EMC performance is attained.
Preparing the Environment
Ensure that the environment in which you are using the NI-9238 meets the following
specifications.
Operating temperature (IEC 60068-2-1, IEC 60068-2-2)
-40 °C to 70 °C
Operating humidity (IEC 60068-2-30)
10% RH to 90% RH, noncondensing
© National Instruments
7
NI-9238 Getting Started
Содержание NI-9238
Страница 1: ...NI 9238 Getting Started 2022 07 06 ...