National Instruments Corp.
19
NI 9237E Operating Instructions and Specifications
modules must share a single master timebase source. Refer to the
software help for information about configuring the master
timebase source for the NI 9237E. Visit
ni.com/info
and enter
cseriesdoc
for information about C Series documentation.
The following equation provides the available data rates of the
NI 9237E:
f
s
=
where
n
is any integer from 1 to 31.
However, the data rate must remain within the appropriate data rate
range. Refer to the
section for more information
about the data rate range. When using the internal master timebase
of 12.8 MHz, the result is data rates of 50 kS/s, 25 kS/s, 16.67 kS/s,
and so on down to 1.613 kS/s, depending on the value of
n
. When
using an external timebase with a frequency other than 12.8 MHz,
the NI 9237E has a different set of data rates.
Note
The NI cRIO-9151 R Series Expansion chassis
does not support sharing timebases between modules.
f
M
256
n
---------------------