NI 9235/9236
Yo
u
can connect a q
u
arter-bridge sensor to each channel. Each
channel has an EXC terminal that provides the excitation voltage
stim
u
l
u
s, an AI terminal that meas
u
res the bridge voltage, and an
RC terminal that provides the q
u
arter-bridge completion. Refer to
Fig
u
re 2 for an ill
u
stration of how to connect q
u
arter-bridge
sensors to the NI 9235/9236.
Figure 2.
Quarter-Bridge Connections for the NI 9235/9236
*
For best system accuracy, set up the connections to EXC and RC with
equal lengths of an identical wire type and gauge.
**
For best system accuracy, set up the connection to AI directly at the
sensor instead of shorting AI to RC directly at the terminals.
EXC
*
AI
**
RC
*
NI 9235/9236