manualshive.com logo in svg
background image

NI 9227 Calibration Procedure

|

© National Instruments

|

5

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9227.

Note

Completing the Accuracy verification procedure verifies both the 5 A

rms

 and 

10 A

rms

 accuracy specifications listed the 

NI 9227 Operating Instructions and 

Specifications

.

1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9227 to the calibrator as shown in Figure 1.

Figure 1.  

Accuracy Verification Connections

3.

Set the calibrator current to a Test Point value indicated in Table 4.

4.

Set the calibrator to Operate mode (OPR).

5.

Acquire and average samples.
a.

Create and configure an AI current task according to Table 2.

b.

Configure the task timing according to Table 3.

c.

Start the task.

Table 2.  

NI 9227 Configuration for Accuracy Verification

Physical Channels

Minimum 

(A

rms

)

Maximum 

(A

rms

)

Terminal 

Configuration

AI0:3

-5.0

5.0

Differential

Table 3.  

NI 9227 Timing Configuration for Accuracy Verification

Sample Mode

Samples Per Channel

Rate (kS/s)

Finite Samples

5000

50

NI 9227

C

a

li

b

r

a

tor

HI

LO

AI1

+

AI0

+

AI2

+

AI

3

+

Содержание NI 9227

Страница 1: ...Matching Verification 6 Phase Matching Verification 8 Adjustment 9 Accuracy Adjustment 9 EEPROM Update 10 Reverification 10 Accuracy Under Calibration Conditions 10 Worldwide Support and Services 11 S...

Страница 2: ...ctions and Specifications NI 9227 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Страница 3: ...device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The device temperature will be greater than the ambient temperature Keep relative humid...

Страница 4: ...em safety ground 4 Install the module in slot 8 of the NI cDAQ 9178 chassis Leave slots 1 through 7 of the NI cDAQ 9178 chassis empty 5 Connect the NI cDAQ 9178 chassis to your host computer 6 Connect...

Страница 5: ...Figure 1 Accuracy Verification Connections 3 Set the calibrator current to a Test Point value indicated in Table 4 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and c...

Страница 6: ...are derived using the values in Table 13 Gain Matching Verification Complete the following procedure to determine the As Found status of the NI 9227 1 Connect the NI 9227 to the function generator as...

Страница 7: ...hannel to channel gain difference to the maximum limit of 0 13 dB 7 Disable the function generator 8 Disconnect the calibrator from the device If the Gain Matching verification procedure determines th...

Страница 8: ...he function generator 5 Acquire samples a Create and configure an AI current task according to Table 9 b Configure the task timing according to Table 10 c Start the task Table 8 Function Generator Con...

Страница 9: ...peration required to adjust the NI 9227 Accuracy Adjustment Complete the following procedure to adjust the accuracy performance of the NI 9227 Note Completing the Accuracy adjustment procedure adjusts...

Страница 10: ...g an external calibration setting the C Series calibration temperature and closing the external calibration Reverification Repeat the Verification section to determine the As Left status of the device...

Страница 11: ...d in Table 4 are derived using the values in Table 13 Note For operational specifications refer to the most recent NI 9227 Operating Instructions and Specifications online at ni com manuals Worldwide...

Отзывы: