NI 9219 Calibration Procedure
Figure 1.
Voltage Accuracy Verification Connections to the NI 9219
2.
Set the source to the appropriate test point value indicated in Table 4. Use the smallest
available range needed to source each test point. Using a range larger than necessary creates
unwanted additional noise.
3.
Configure the DMM for a voltage measurement in the lowest appropriate range according
to the test point from Table 4.
4.
Enable Auto Zero on the DMM.
5.
Acquire a voltage reading with the DMM. Record this measurement as V
ref
.
6.
Acquire and average samples.
a.
Create an AI voltage task on the NI 9219 according to Table 2.
b.
Configure the AI voltage task timing according to Table 3.
Table 2.
NI 9219 Configuration for Voltage Accuracy Verification
Measurement Type
Minimum (V)
Maximum (V)
Voltage
–60
60
–15
15
–4
4
–1
1
–0.125
0.125
V+
V–
HI
V
LO
NI 9219
DMM
S
ource
HI (4)
LO (5)
Ch 0
HI (4)
LO (5)
Ch 1
HI (4)
LO (5)
Ch 2
HI (4)
LO (5)
Ch 3