NI 9219 Calibration Procedure
|
© National Instruments Corporation
|
25
Figure 13.
Half-Bridge Accuracy Verification Connections to the NI 9219 (Configuration 4)
7.
Acquire and average samples.
a.
Configure DMM1 for a voltage measurement in the 10 V range.
b.
Enable Auto Zero on DMM1.
c.
Configure DMM2 for a voltage measurement in the 10 V range.
d.
Enable Auto Zero on DMM2.
e.
Create an AI bridge (V/V) task on the NI 9219 to Table 26.
f.
Configure the AI bridge (V/V) task timing according to Table 29.
g.
Start the task.
h.
Acquire voltage measurements with both DMMs.
i.
Record the DMM1 measurements as V
DMM1
.
Table 28.
NI 9219 Configuration for Half-Bridge Accuracy Verification
Measurement
Type
Min (V)
Max (V)
Excitation
Source
Excitation
Value
Bridge
Config
Half-Bridge
500 mV/V
–0.5
0.5
Internal
2.5
Half-Bridge
Table 29.
NI 9219 Timing Configuration
ADC Timing Mode
Sample Timing Type
High Resolution
On Demand
HI
V
LO
DMM2
NI 9219
HI
V
LO
DMM1
V+
V–
S
ource
EX+ (3)
HI (4)
EX– (5)
Ch
x
1 k
Ω
± 1%
1 k
Ω
± 1%
5 k
Ω
± 1%
5 k
Ω
± 1%