NI 9214 Operating Instructions and Specifications
16
ni.com
optimizes sample rate and signal bandwidth. Refer to the
section for more information abo
u
t the high-speed
and high-resol
u
tion timing modes. Refer to the software help for
information abo
u
t setting the timing mode, or conversion time,
in software. Visit
ni.com/info
and enter
c
s
erie
s
doc
for
information abo
u
t C Series doc
u
mentation.
Temperature Measurement Accuracy
Considerations
Temperat
u
re meas
u
rement errors depend partly on the
thermoco
u
ple type, the acc
u
racy of the thermoco
u
ple, the
temperat
u
re being meas
u
red, the resistance of the thermoco
u
ple
wires, and the cold-j
u
nction temperat
u
re. Refer to the
section in the
section for the
mod
u
le acc
u
racy for each thermoco
u
ple type when connected to
the NI 9214.
For the best acc
u
racy res
u
lts, keep temperat
u
re gradients across
NI 9214 terminals to a minim
u
m, n
u
ll the lead wire resistance, and
enable the a
u
tozero channel. Refer to the
section, the
section,
and the
section for more information.
NI_9214.fm Page 16 Monday, November 1, 2010 9:52 AM